Introduction
Posterior screw–rod fixation has become the preferred method of atlantoaxial fixation with the advent of C1 lateral mass screws [
1,
2]. In addition, the emergence of multiple techniques of posterior screw fixation for C1 and C2 has further expanded the versatility of posterior atlantoaxial fixation, allowing greater flexibility in screw configuration and expanding its clinical applicability [
3,
4]. Reduction and stabilization represent the primary surgical objectives for atlantoaxial dislocation (AAD) [
5]. In posterior screw–rod fixation, bilateral screw placement in C1 and C2 offers procedural efficiency. The posterior atlantoaxial screw–rod (PASR) fixation system provides not only rigid stabilization but also intrinsic reduction capacity through mechanical distraction and compression.
At present, more attention is being paid to the research of screw fixation techniques for the improvement of the PASR fixation system [
6–
9], while less attention is being paid to the improvement of instruments. Given the maturation of atlantoaxial screw fixation techniques, current scientific efforts should be redirected to the development of instrumentation that facilitates a more simplified and effective surgical reduction of AAD. Posterior C1–C2 screw–rod fixation is currently the standard approach for atlantoaxial stabilization. Nevertheless, significant limitations persist when applied to AAD reduction surgery. First, an inadequate reduction force frequently leads to intraoperative reduction failure. Second, prebending deformation causes rod rotation within screw heads, compromising directional stability and weakening the reduction effect. Third, in the case of severe dislocation, the gap between C1 and C2 is quite large, which makes it difficult to insert the rod into the screw groove. To address these shortcomings, we designed a novel PASR fixation system. This study compared surgical results between our modified PASR fixation system and a conventional PASR fixation system and evaluated the preliminary clinical efficacy of this novel system. To our knowledge, this is the first study to report the instrument improvement of the PASR fixation system for AAD treatment.
Materials and Methods
Ethics statement
This study was approved by the Institutional Review Board (IRB) of the General Hospital of Southern Theater Command ((IRB approval no., NZLLKZ2025256).
Study design and patient selection
From June 2016 to June 2021, a total of 39 consecutive patients with AAD were recruited at the General Hospital of Southern Theater Command. Irreducible AAD is defined as a dislocation between the atlas (C1) and axis (C2) that persists despite flexion–extension dynamic radiographs and a trial of in-line skull traction.
All participants underwent PASR fixation. They were randomized into two treatment groups using stratified randomization that accounted for age (≥60 years vs.<60 years), sex, and dislocation type (reducible vs. irreducible). Within each stratum, randomization sequences were generated in a 1:1 allocation ratio through IBM SPSS ver. 19.0 (IBM Corp., Armonk, NY, USA) by an independent researcher not involved in surgical procedures or data collection. Of the 39 patients, 18 were treated with the novel PASR fixation system (novel PASR group) and 21 were treated with the conventional PASR fixation system (conventional PASR group). All surgeries were performed by the same surgeon (X.Y.M.). Demographic characteristics (age and sex), type of dislocation (reducible or irreducible), disease duration, and follow-up time were compared between the groups.
Introduction of the novel PASR fixation system
Our novel PASR fixation system included C1 reduction screws, C2 fulcrum screws, and antirotation rods (WEGO, Weihai, China). The C1 reduction screw, also known as a long-arm screw, was designed in reference to a thoracolumbar long-arm screw and had two types of polyaxial and uniaxial screws (
Fig. 1A, B). The C2 fulcrum screw was designed to increase the bottom of the screw groove by 2 or 4 mm, and it had polyaxial and uniaxial screws (
Fig. 1A, B). One end of the antirotation rod was designed as a “D-shaped” structure about 15 mm long, which was inserted into the C2 screw groove for automatic antirotation, and the other end was a cylindrical structure about 35 mm long, which could be cut according to the required length during the operation (
Fig. 1A, B). Comparison of reduction performance between the conventional and novel PASR fixation systems is presented in
Fig. 1C and D. The operation procedure and reduction principle of the new instruments are shown in
Fig. 2A–G.
Surgical procedure
After general anesthesia, the patient was placed in the prone position with moderate cervical flexion. Skull traction (4 kg) was applied to stabilize the head. A longitudinal incision of about 6–8 cm was made in the posterior middle of the neck from the occipital to the C2 spinous process. The occipital bone, the posterior arch of C1, and the C2 lateral mass were exposed by subperiosteal dissection, and the cervical semispinalis cervicis muscle attached to the C2 spinous process was retained.
A pedicle screw or a lateral mass screw was used for C1 fixation, with a pedicle screw being preferred. If the C1 posterior arch was too thin and less than 2.0 mm in height, a C1 lateral mass screw was inserted according to the technique described by Goel and Laheri [
1] and Harms et al. [
2]; otherwise, a C1 pedicle screw was used according to the technique described by Ma et al. [
10]. The newly designed C1 reduction screws were used in place of all C1 screws in the novel PASR group, and conventional C1 screws were used in the conventional PASR group. In the novel and conventional PASR groups, reduction was typically achieved by sequentially tightening nuts along a prebending rod. The adequacy of reduction was then verified using intraoperative X-ray imaging.
A C2 pedicle screw was preferred for C2 fixation. If anatomical variation made it impossible to safely implant C2 pedicle screws, C2 pars screws or C2 translaminar screws were used instead [
11,
12]. The newly designed C2 fulcrum screws were used in place of all C2 screws in the novel PASR group, and C2 fulcrum screws with a bottom height of 2 or 4 mm were selected according to the degree of dislocation in preoperative imaging. For severe dislocation, 4-mm screws were selected to provide greater reduction force, and for mild-to-moderate dislocation, 2-mm screws were used. Conventional C2 screws were used in the conventional PASR group.
After two antirotation rods were trimmed, the “D-shaped” end was inserted into the C2 fulcrum screw groove, and the cylindrical part was placed into the C1 reduction screw groove (
Fig. 2A–E). The nuts of the C2 fulcrum screws were tightened, and the direction of the rods was automatically fixed. Then, the nuts of the C1 reduction screws were gradually tightened to complete the reduction (
Fig. 2B, F, G). After good reduction was confirmed by intraoperative fluoroscopy, the long arms of the C1 reduction screws were broken (
Fig. 2C). In the conventional PASR group, two prebending conventional cylindrical rods were used. Subsequently, autografts taken from the iliac crest were grafted.
Data collection and postoperative outcomes
The surgical time and estimated blood loss (EBL) were recorded. The atlantodental interval (ADI) and atlantoaxial angle (AAA) were measured before and after the surgery to assess the reduction. Japanese Orthopaedic Association (JOA) scores were used to assess neurological function. The Visual Analog Scale (VAS) was used to assess the degree of occipital neck pain. The Neck Disability Index (NDI) was used to evaluate cervical spine function.
Patients were followed up at 3, 6, 9, and 12 months, and then once per year or as needed. Surgical complications were documented after the surgery and during the follow-up. Cervical X-ray imaging and computed tomography (CT) scans were performed at each follow-up for the assessment of device failure, stability, and bone fusion.
Statistical analysis
Data were analyzed statistically with the Mann-Whitney U test, using IBM SPSS 19.0 software (IBM Corp.). The significance threshold was set at p<0.05.
Results
Patient demographics
Of the 39 patients, 18 were enrolled in the novel PASR group and 21 were enrolled in the conventional PASR group. There were no significant differences between groups with respect to age (
p=0.603), gender (
p=0.433), type of dislocation (
p=0.197), disease duration (
p=0.132), and follow-up time (
p=0.124). The clinical and demographical characteristics for all participants are presented in
Table 1.
Surgical outcomes
The surgery was performed successfully in all 39 patients. The surgical time was significantly lower in the novel PASR group than in the conventional PASR group (
p<0.01), and there was no significant difference in EBL between the groups (
p>0.05) (
Table 2). No vertebral artery or spinal cord injury occurred during the operations in both groups.
Clinical outcomes
There were no differences in preoperative data between the two groups (
p>0.05). In both groups, the ADI, AAA, JOA score, VAS score, and NDI were significantly better at the last follow-up assessment than at the preoperative assessment (
p<0.01). There were no significant differences in the AAA, VAS score, and NDI at the final follow-up between the two groups (
p>0.05). However, at the final follow-up, the ADI was significantly lower in the novel PASR group than in the conventional PASR group (
p<0.05), while the JOA score was significantly higher in the novel PASR group than in the conventional PASR group (
p<0.05). All patients achieved bone fusion, which was confirmed on CT scans, and there was no difference in fusion time between the two groups (
p>0.05) (
Table 2). No instrument loosening or breakage, or loss of reduction was noted on X-ray imaging and CT scans during the follow-up (
Figs. 3–
5).
Discussion
This study evaluated the clinical outcomes of a novel PASR fixation system and compared it with the conventional PASR fixation system in the treatment of AAD. All patients treated with the novel system achieved stable fixation, sustained radiological correction, and better clinical benefit, indicating its safety and effectiveness in the treatment of AAD.
Our findings showed that the novel system significantly shortened the operative time, which was attributable to its more efficient reduction process and improved reduction capability. Although there was no difference in the AAA between the two study groups at the last follow-up, the ADI was lower in the novel PASR group than in the conventional PASR group, indicating that this novel PASR fixation system has a superior reduction capacity and helps to relieve cord compression. This may also explain why the JOA score at the final follow-up was higher in the novel PASR group than in the conventional PASR group. Furthermore, successful operative treatment of two cases with irreducible AAD without anterior or posterior soft tissue release by only using this novel PASR fixation system suggests that this new approach has a great reduction ability.
The advantages of the novel PASR fixation system are associated with several key design elements. The C1 reduction screw, with an extended arm, allows gradual sliding and reduction during nut tightening, avoiding the need for additional tools. The C2 multiaxial fulcrum screw and monoaxial fulcrum screw, designed with a raised groove (2 or 4 mm), increase the vertical difference between the C1 and C2 screws, providing greater leverage for reduction (
Fig. 1B). It is proposed that the reduction ability of a monoaxial screw is greater than that of a multiaxial screw. Therefore, for severe dislocation, a monoaxial fulcrum screw would be preferred. In addition, the antirotation rod with a “D-shaped” end prevents rod rotation during tightening, eliminating the need for auxiliary instruments in a narrow surgical field and reducing the risk of spinal cord injury. The cylindrical design of the other end of the rod makes it easy to place the rod into the groove of the atlas screw, and it can be automatically attached and anastomosed. Collectively, these features enable simultaneous reduction and fixation, which can simplify the surgical procedure and improve reduction efficiency compared with conventional systems.
Compared with earlier fixation techniques, the novel PASR fixation system offers additional benefits. The Magerl screw technique, once the gold standard, provides strong biomechanical stability but requires preoperative reduction, has limited applicability in patients with a short neck or obesity, and carries the risks of vertebral artery injury and radiation exposure [
13–
15]. Conventional PASR fixation systems partially address these issues [
16], but they have not been specifically designed for the unique anatomical and biomechanical challenges of the atlantoaxial region. Conventional posterior cervical screws are all short-arm screws [
17]. When the anterior dislocation of C1 is severe, it is very inconvenient and dangerous to use the approach involving a lifting screw and press rod to assist the operation in a narrow space. The C2 screw is the reduction fulcrum in the PASR fixation system, and theoretically, a higher fulcrum is associated with a better reduction effect [
18,
19]. To enhance the reduction capability, Zhou et al. [
8] used pars screws with the entry point placed as inferiorly as possible to increase the height difference between the C1 and C2 screws. Clinical applications have shown that the reduction effect of the pars screw is better; however, due to the shorter length of the channel, it may lack sufficient compressive strength when reduction resistance is excessive or in the presence of osteoporosis. The novel PASR fixation system builds on these foundations and introduces structural innovations tailored to the treatment of AAD, thereby enhancing safety, efficiency, and reduction success rates. The system has the ability to achieve automatic reduction through the tightening of nuts into sliding C1 screw heads (the C1 screws have long heads), the presence of C2 screws with a higher abutment level, and the presence of antirotation rods.
This study has some limitations. The study was retrospective and involved a relatively small sample size, which may limit the generalizability of the results. Furthermore, biomechanical testing and longer-term follow-up were not performed. Future studies should include larger prospective cohorts and biomechanical validation to further confirm the advantages of this system and clarify its indications.
In summary, the novel PASR fixation system provides a safe and effective alternative to conventional fixation for AAD. Its ability to achieve automatic reduction during fixation shortens the operative time, enhances reduction, and may improve neurological recovery, particularly in cases of irreducible AAD.
Conclusions
We developed a novel PASR fixation system to treat AAD. Compared with the conventional PASR fixation system, the novel fixation system could better simplify the atlantoaxial fixation procedure and achieve atlantoaxial dislocation reduction without additional surgical release.