Zhong, Tan, Ma, Yuan, Wen, Zhou, Xiong, Yu, and Cao: Biomechanical comparison of multirod constructs for severe thoracolumbar Pott’s deformity following posterior vertebral column resection: a finite element study

Abstract

Study Design

Finite element study.

Purpose

This study aimed to compare the biomechanical performance of three multirod constructs used in posterior vertebral column resection (PVCR) for severe thoracolumbar Pott’s deformity.

Overview of Literature

The optimal configuration of the multirod constructs for spinal deformity correction remains controversial.

Methods

A patient-specific three-dimensional spinal model was constructed, and PVCR was simulated using modeling software. Three fixation configurations were evaluated: (A) conventional dual rods with accessory rods bridging the osteotomy site, (B) short-segment fixation adjacent to the osteotomy with long proximal and distal rods, and (C) long rods engaged with staggered screws. A 10-N·m moment was applied to simulate flexion, extension, lateral bending, and axial rotation. Implant stress and vertebral displacement were recorded for each model.

Results

Under flexion, the maximum implant stresses of Models A, B, and C were 92, 99, and 131.1 MPa, respectively, occurring at the most cranial screws. Under extension, the corresponding values were 95, 101, and 131.6 MPa, respectively, occurring at the most cranial screws. Under lateral bending, the maximum stresses were 125.9, 118.5, and 155 MPa, respectively, at the cranial rod segments. Under axial rotation, the corresponding values were 213.7, 199, and 214 MPa, respectively, at the most cranial screws. The respective maximum vertebral displacements for Models A, B, and C were 2.4, 2.5, and 2.8 mm in flexion; 2.7, 2.8, and 3.1 mm in extension; 1.3, 1.4, and 2.3 mm in lateral bending; and 1.6, 2.4, and 2.1 mm in rotation.

Conclusions

Implant stress in PVCR was predominantly concentrated at the most cranial screws and cranial rod segments. In this finite element analysis, Models A and B demonstrated lower implant stress and construct displacement than Model C under most loading conditions.

Key Points
  • Posterior vertebral column resection is an effective surgical technique to correct severe thoracolumbar Pott’s deformity.

  • Multirod constructs have been used to reduce the risk of implant-related complications in complex spinal deformity surgery.

  • In this finite element study, Models A and B showed lower implant stress and construct displacement than Model C under most loading conditions.

  • These findings provide biomechanical insights into the load-sharing characteristics of different multirod constructs; however, further experimental and clinical validation is required.

Introduction

Spinal deformities are characterized by abnormal curvature and misalignment of the spinal column, which may result in chronic back pain, respiratory dysfunction, neurological deficits, and reduced quality of life [1]. Three-column osteotomies are commonly performed to correct moderate to severe spinal deformities. Compared with Schwab grade 3 and 4 osteotomies, posterior vertebral column resection (PVCR), also classified as a Schwab grade 5 osteotomy, involves a more extensive bone resection, greater local release, and larger capacity for deformity correction [2]. However, the substantial corrective potential of PVCR is associated with increased instability at the osteotomy site. Inadequate internal fixation or insufficient anterior column support may compromise correction and increase the risk of implant-related complications. Rod fracture, screw loosening, and pseudarthrosis not only affect patient outcomes but also increase the need for revision surgeries [3]. Previous studies found that the incidence rate of rod fracture and late complications, such as pseudarthrosis, related to excessive stress on the rod-screw construct ranges from 15.8% to 22.0% [4-8].
Previous clinical analyses suggested that the use of multirod constructs may reduce the risk of implant failure following correction surgery for severe thoracolumbar Pott’s kyphosis [9]. Several multirod construct configurations have been described in clinical practice; however, biomechanical investigations comparing these configurations remain limited [10,11]. Consequently, the biomechanical characteristics of different multirod constructs and their relative advantages remain unclear.
Therefore, the present study developed three-dimensional (3D) finite element models representing three commonly used multirod constructs in PVCR for thoracolumbar Pott’s deformity. Finite element analysis was performed to evaluate implant stress and vertebral displacement under different loading conditions. This study aimed to compare the biomechanical behavior of these configurations and provide biomechanical insights into their load-sharing characteristics.

Materials and Methods

Case selection

A 16-year-old male patient with severe thoracolumbar Pott’s kyphosis who underwent PVCR at The First Affiliated Hospital of Nanchang University was selected for this study. The lesion involved the T9–T11 vertebrae. The preoperative Konstam angle, defined as the angle between the superior and inferior endplates of the normal vertebrae adjacent to the lesion, was 111.6° [12]. Institutional Review Board approval was not required at The First Affiliated Hospital of Nanchang University for this type of computational study that used anonymized imaging data. Informed consent was waived due to the retrospective nature of the study and use of deidentified data.

Multiple-rod construct models

Based on previously reported surgical techniques and clinical application of multirod constructs in Pott’s deformity correction, three representative fixation configurations were selected for the analysis [10,11,13,14]: Model A, conventional dual rods supplemented with two accessory short rods bridging the osteotomy site and connected using domino connectors; Model B, short-segment screw–rod fixation at the two vertebrae adjacent to the osteotomy site combined with long proximal and distal rods; and Model C, bilateral long rods engaged with staggered pedicle screws placed in the proximal three and distal three vertebrae. A schematic illustration of the three multirod constructs is presented in Fig. 1.

Finite element model construction

Thin-slice computed tomography images of the patient’s thoracolumbar spine (slice thickness: 1 mm) were imported in DICOM format into Mimics (Materialise, Plymouth, MI, USA) to reconstruct the spinal geometry (Fig. 2). Using grayscale thresholding, region growing, and segmentation tools, the vertebrae from T6 to L2 were isolated, and a preliminary 3D model was generated (Fig. 3A).
Then, the reconstructed model was exported to 3-matic software for geometric processing and surgical simulation. To simulate PVCR, the diseased vertebrae (T9–T11) were removed, leaving three vertebrae above and below the osteotomy level. The vertebral structures were further processed through mesh repair, smoothing, and surface wrapping to obtain a clean geometric model. Cortical and cancellous bone regions were generated using offset and Boolean operations. Subsequently, intervertebral discs, articular cartilage, and spinal ligaments were added to complete the spinal structural model. The model was further smoothed using Geomagic Studio ver. 12.0 (Raindrop Geomagic Inc., Morrisville, NC, USA). Then, pedicle screws, rods, and titanium mesh cages were constructed and assembled using Unigraphics NX ver. 12.0 (Siemens PLM Software, Munich, Germany) according to the predefined fixation configurations. The final solid model was exported in STP format for finite element analysis.

Model setup and boundary conditions

The postoperative spinal model was imported into ANSYS Workbench ver. 19.0 (ANSYS Inc., Canonsburg, PA, USA) for mesh generation, material property assignment, contact definition, and loading conditions. The intertransverse, interspinous, supraspinous, capsular, and ligamentum flavum were simulated using linear spring elements attached at their anatomical insertion sites. Bonded contact conditions were applied between the cortical and cancellous bones, vertebral bodies and intervertebral discs, and ligaments and vertebrae. Articular cartilages were modeled using frictionless contact.
The models were meshed using tetrahedral elements. The mesh size was controlled within 3 mm for most structures, with finer meshes applied to intervertebral discs (1 mm), cartilage (0.5 mm), and implants (1.5 mm). There were 467,253 elements and 124,675 nodes in Model A; 491,062 elements and 126,441 nodes in Model B; and 443,221 elements and 119,556 nodes in Model C. All materials were assumed to be linear elastic and isotropic. The material properties of bone, discs, cartilage, and implants were assigned based on previous studies, as shown in Table 1.

Loading conditions and outcome measures

A Cartesian coordinate system was established for the spinal model (Fig. 3B). The inferior surface of the L2 vertebra was fully constrained, restricting all six degrees of freedom to simulate fixation. A 10-N·m moment was applied to the superior surface of the T6 vertebra to simulate four physiological loading conditions: flexion, extension, lateral bending, and axial rotation. Vertebral displacement was defined as the maximum nodal translational displacement of the vertebrae relative to the fixed inferior surface of L2 following load application. The maximum displacement typically occurred at the anterior superior edge of the most cranial vertebra or superior articular process (Fig. 3A), depending on the loading condition, and was used to represent the overall deformation of the spinal construct. The implant stress of the internal fixation system was also recorded for each model.

Results

Internal fixation stress

Under flexion, the maximum implant stresses in Models A, B, and C were 92, 99, and 131.1 MPa, respectively, occurring at the most cranial screws (Fig. 4A). Under extension, the corresponding maximum stresses were 95, 101, and 131.6 MPa, respectively, also occurring at the most cranial screws (Fig. 4B). Under lateral bending, the maximum stresses in Models A, B, and C were 125.9, 118.5, and 155 MPa, respectively, which are primarily concentrated at the cranial rod segments (Fig. 4C). Under axial rotation, the maximum stresses in Models A, B, and C were 213.7, 199, and 214 MPa, respectively, again occurring at the most cranial screws (Fig. 4D). The maximum implant stress values and corresponding percentage differences among the three models are shown in Table 2.

Vertebral displacement

Under flexion, the maximum vertebral displacements in Models A, B, and C were 2.4, 2.5, and 2.8 mm, respectively, occurring at the anterior superior edge of the most cranial vertebra (Fig. 5A). Under extension, the corresponding displacements were 2.7, 2.8, and 3.1 mm, respectively, also occurring at the anterior superior edge of the most cranial vertebra (Fig. 5B). Under lateral bending, the maximum displacements in Models A and B were 1.3 and 1.4 mm, respectively, occurring at the anterior superior edge of the most cranial vertebra. In Model C, the maximum displacement of 2.3 mm occurred at the superior articular process of the most cranial vertebra (Fig. 5C). Under axial rotation, the maximum displacement in Model A was 1.6 mm at the anterior superior edge of the most cranial vertebra, whereas the maximum displacements in Models B and C were 2.4 and 2.1 mm, respectively, which occurred at the superior articular process of the most cranial vertebra (Fig. 5D). The maximum displacement values and percentage differences among the three models are presented in Table 3, while the overall stress and displacement characteristics are shown in Table 4.

Discussion

Implant-related complications remain an important problem for surgeons performing PVCR for severe spinal deformities [15,16]. Previous studies identified several factors associated with implant failure after spinal deformity correction, including patient age, body mass index, preoperative sagittal balance, fusion length, correction magnitude, fixation strategy, and implant material [4,7]. Among these factors, the application of multirod constructs reduced rod stress and decreased the risk of mechanical complications [4,7,10,17-19]. However, because several multirod configurations are used in clinical practice and the number of patients receiving each construct is relatively small, it is difficult to determine the relative advantages of different configurations through clinical studies alone [13]. In the present study, three commonly used multirod constructs were analyzed using finite element analysis to compare implant stress and vertebral displacement under different loading conditions. Across all three models, the highest implant stresses were consistently observed at the most cranial screws or adjacent rod segments, suggesting that these regions represent potential mechanical weak points in the fixation system. Under flexion, extension, and lateral bending conditions, Models A and B generally exhibited lower implant stresses and smaller vertebral displacements than Model C, indicating improved load sharing across the fixation construct. These differences can be explained by the structural characteristics of the constructs. In Model A, the vertebrae are connected through a conventional dual-rod system supplemented with accessory rods spanning the osteotomy site. The additional rods increase the structural stiffness of the construct and distribute loads across multiple rods, thereby reducing stress concentration. Conversely, in Model C, staggered screws with two long rods that are not simultaneously connected to all vertebrae were used. This configuration may result in less uniform load transfer and greater localized stress concentration, which may explain the relatively higher implant stresses and larger displacements observed in this model.
Several classification systems for multirod constructs have been proposed. El Dafrawy et al. [11] categorized multirod systems into primary and secondary rods, in which the primary rods provide the main structural support and secondary rods act as supplemental elements to enhance stability. Subsequently, Ramey et al. [10] classified rods into primary, secondary, satellite, accessory, kickstand, and delta rods based on their mechanical roles and positions within the construct. According to this classification, the short rods used in Model A correspond to accessory rods connected to the primary rods through domino connectors, whereas the short rods used in Model B function as satellite rods that span the osteotomy level. In the present study, Models A and B demonstrated different stress distribution patterns depending on loading conditions. Model A showed slightly lower stresses under flexion and extension, whereas Model B exhibited relatively lower stresses under lateral bending and axial rotation. These findings suggest that different multirod configurations may influence load-sharing characteristics in different ways. Previous biomechanical studies demonstrated that the addition of satellite rods across osteotomy levels can reduce stresses on primary rods and improve construct durability [3,7,20,21]. For example, Shekouhi et al. [3] found that multirod constructs incorporating satellite rods in lumbar pedicle subtraction osteotomy significantly reduced stresses on primary rods compared with conventional dual-rod constructs. These findings are generally consistent with the concept that increasing the number of rods may enhance load distribution across the instrumentation system.
From a surgical perspective, adequate space for the placement of accessory or satellite rods is necessary when constructing multirod systems. The use of domino connectors to attach accessory rods, as in Model A, may require partial removal of posterior vertebral components to accommodate the rods. In satellite rod constructs, such as Model B, screw trajectories and rod alignment must be carefully planned to prevent interference between adjacent rods. Conversely, staggered screw constructs similar to Model C may present technical challenges because differences in pedicle orientation across spinal regions can complicate rod placement. These anatomical and technical considerations may influence the feasibility of different multirod configurations during surgery. Notably, under the simulated conditions of this finite element analysis, Model A demonstrated the most favorable displacement profile among the tested constructs; however, these findings do not justify prescriptive clinical recommendations without further validation. Conversely, Model C showed less favorable results in this finite element model under specific loading conditions and thus should not be interpreted as evidence against its clinical use without additional experimental or clinical validation. Another notable finding of this study is that the highest implant stresses consistently occurred at the most cranial screws or adjacent rod segments. This observation suggests that the upper instrumented vertebra may represent a mechanically vulnerable region within the fixation system. Therefore, from a biomechanical perspective, adequate fixation strength at the cranial end of the construct may be important in maintaining overall construct stability.
Several limitations of this study should be acknowledged: First, the finite element model was derived from imaging data of a single patient, which may limit the generalizability of the findings to broader patient populations. Second, only four simplified loading conditions were simulated, whereas in vivo spinal loading is considerably more complex. Third, the finite element model assumed linear elastic material properties and did not fully account for variations in bone density or nonlinear soft tissue behavior. Lastly, experimental validation using cadaveric models was not performed. Therefore, the results of this study should be interpreted as model-specific biomechanical observations, and further experimental and clinical investigations are required to confirm these findings.

Conclusions

In this finite element analysis, Model A generally demonstrated lower construct displacement under the tested loading conditions compared with the other configurations. However, these findings should be interpreted as relative, model-specific comparisons rather than definitive evidence of clinical superiority. The clinical significance of these differences (e.g., with respect to implant yield/fatigue thresholds, safety margins, or patient-centered outcomes) cannot be directly inferred from the present simulation alone. Therefore, construct selection should be framed as a goal-dependent decision that may vary based on surgical objectives (e.g., prioritizing rotational stress minimization versus global displacement control) and anatomic feasibility. Future experimental and clinical validation is warranted before any clinical recommendation can be made.

Notes

Conflict of Interest

No potential conflict of interest relevant to this article was reported.

Funding

This work was supported by the National Key R&D Program of China (No. 2023YFC2507700), National Natural Science Foundation of China (No. 82360419), Double-Thousand Plan Program of Jiangxi Province (No. 8210298001), Clinical Cultivation Project of The First Affiliated Hospital of Nanchang University (No. YFYLCYJPY202202), Jiangxi Provincial Natural Science Foundation (No. 20232BAB216033), and Project of Jiangxi Provincial Administration of Traditional Chinese Medicine (No. 2023A0282).

Author Contributions

Conceptualization: KC, JZ, SM, PW. Data curation: KC, JZ, WT, SM, ZY, JX, HY. Formal analysis: KC, JZ, WT, SM, ZZ, JX. Funding acquisition: KC, JZ, ZZ. Methodology: KC, JZ, WT, SM, PW, ZZ, JX. Project administration: KC, SM, PW. Visualization: WT, ZZ, JX, HY. Writing–original draft: JZ, WT, SM. Writing–review & editing: ZY, PW, ZZ, JX, HY, KC. Final approval of the manuscript: all authors.

Fig. 1.
The three multiple-rod construct models.
asj-2025-0614f1.jpg
Fig. 2.
(A–D) Patient computed tomography data processing in Mimics software.
asj-2025-0614f2.jpg
Fig. 3.
The three-dimensional model of the target vertebra (A) and its vertical positioning along the x, y, and z axes (B). The displacement measurement points are illustrated with three green markers (A).
asj-2025-0614f3.jpg
Fig. 4.
The internal fixation stress of each model in different conditions. (A) Flexion. (B) Extension. (C) Lateral bending. (D) Rotation.
asj-2025-0614f4.jpg
Fig. 5.
The displacement of each model in different conditions. (A) Flexion. (B) Extension. (C) Lateral bending. (D) Rotation.
asj-2025-0614f5.jpg
Table 1.
Material properties used in the model
Component Elastic modulus (MPa) Poisson’s ratio Stiffness (N/mm)
Cortical bone 12,000 0.30 -
Cancellous bone 100 0.30 -
Facet cartilage 35 0.40 -
Intervertebral disc 4.20 0.45 -
Screw/titanium cage/rod 110,000 0.36 -
Anterior longitudinal ligament 20 0.30 -
Posterior longitudinal ligament 70 0.30 -
Ligamentum flavum - - 19,201
Interspinous ligament - - 14,598
Supraspinous ligament - - 4,298
Intertransverse ligament - - 2,407
Capsular ligament - - 49,809

N/mm, Newton/millimeter.

Table 2.
Maximum internal fixation stress for each model
Model Maximum internal fixation stress (MPa, %)
Flexion Extension Lateral bending Axial rotation
A 92.0 (100.0) 95.0 (100.0) 125.9 (100.0) 213.7 (100.0)
B 99.0 (107.6) 101.0 (106.3) 118.5 (94.1) 199.0 (93.1)
C 131.1 (142.5) 131.6 (138.5) 155.0 (123.1) 214.0 (100.1)

Percentages in parentheses are relative to Model A.

Table 3.
Maximum displacement for each model
Model Maximum displacement (mm, %)
Flexion Extension Lateral bending Axial rotation
A 2.4 (100.0) 2.7 (100.0) 1.3 (100.0) 1.6 (100.0)
B 2.5 (104.2) 2.8 (103.7) 1.4 (107.7) 2.4 (150.0)
C 2.8 (116.7) 3.1 (114.8) 2.3 (176.9) 2.1 (131.3)

Percentages in parentheses are relative to Model A.

Table 4.
The stress and displacement characteristics for each model
Model Category Flexion Extension Lateral bending Axial rotation
A Stress Low Low - -
Displacement Small Small Small Small
B Stress - - Low Low
Displacement - - Large
C Stress High High High High
Displacement Large Large Large -

References

1. Ames CP, Scheer JK, Lafage V, et al. Adult spinal deformity: epidemiology, health impact, evaluation, and management. Spine Deform 2016;4:310–22. https://doi.org/10.1016/j.jspd.2015.12.009
crossref pmid pmc
2. Schwab F, Blondel B, Chay E, et al. The comprehensive anatomical spinal osteotomy classification. Neurosurgery 2015;76 Suppl 1:S33–41. https://doi.org/10.1227/01.neu.0000462076.73701.09
crossref pmid
3. Shekouhi N, Vosoughi AS, Zavatsky JM, Goel VK, Theologis AA. Biomechanical comparison of multi-rod constructs by satellite rod configurations (in-line vs. lateral) and screw types (monoaxial vs. polyaxial) spanning a lumbar pedicle subtraction osteotomy (PSO): is there an optimal configuration? Eur Spine J 2022;31:3050–9. https://doi.org/10.1007/s00586-022-07331-7
crossref pmid
4. Berjano P, Bassani R, Casero G, Sinigaglia A, Cecchinato R, Lamartina C. Failures and revisions in surgery for sagittal imbalance: analysis of factors influencing failure. Eur Spine J 2013;22 Suppl 6:S853–8. https://doi.org/10.1007/s00586-013-3024-x
crossref pmid pmc
5. Bridwell KH, Lewis SJ, Lenke LG, Baldus C, Blanke K. Pedicle subtraction osteotomy for the treatment of fixed sagittal imbalance. J Bone Joint Surg Am 2003;85:454–63. https://doi.org/10.2106/00004623-200303000-00009
crossref pmid
6. Smith JS, Shaffrey CI, Ames CP, et al. Assessment of symptomatic rod fracture after posterior instrumented fusion for adult spinal deformity. Neurosurgery 2012;71:862–7. https://doi.org/10.1227/NEU.0b013e3182672aab
crossref pmid
7. Smith JS, Shaffrey E, Klineberg E, et al. Prospective multicenter assessment of risk factors for rod fracture following surgery for adult spinal deformity. J Neurosurg Spine 2014;21:994–1003. https://doi.org/10.3171/2014.9.SPINE131176
crossref pmid
8. Barton C, Noshchenko A, Patel V, Cain C, Kleck C, Burger E. Risk factors for rod fracture after posterior correction of adult spinal deformity with osteotomy: a retrospective case-series. Scoliosis 2015;10:30. https://doi.org/10.1186/s13013-015-0056-5
crossref pmid pmc
9. Zhong J, Gao J, Xiong J, et al. What factors were associated with implant failure after posterior vertebral column resection for severe thoracolumbar Pott deformity? Clin Orthop Relat Res 2026;484:989–97. https://doi.org/10.1097/CORR.0000000000003772
crossref pmid
10. Ramey WL, Jack AS, Chapman JR. The lexicon of multirod constructs in adult spinal deformity: a concise description of when, why, and how. J Neurosurg Spine 2021;36:1023–9. https://doi.org/10.3171/2021.10.SPINE21745
crossref pmid
11. El Dafrawy MH, Adogwa O, Wegner AM, et al. Comprehensive classification system for multirod constructs across threecolumn osteotomies: a reliability study. J Neurosurg Spine 2020;34:103–9. https://doi.org/10.3171/2020.6.SPINE20678
crossref pmid
12. Rajasekaran S, Shanmugasundaram TK. Prediction of the angle of gibbus deformity in tuberculosis of the spine. J Bone Joint Surg Am 1987;69:503–9.
crossref pmid
13. Krishnan A, Raj A, Meena U, et al. RCC (reinforced crisscross construct): an easy and effective multi-rod thoracolumbar posterior reconstruction technique. Spine Deform 2022;10:1203–8. https://doi.org/10.1007/s43390-022-00504-w
crossref pmid
14. Sardi JP, Berlin C, Buell TJ, et al. Use of supplemental rod constructs in adult spinal deformity surgery: a review. Oper Neurosurg 2025;29:323–34. https://doi.org/10.1227/ons.0000000000001484
crossref
15. Kavadi N, Tallarico RA, Lavelle WF. Analysis of instrumentation failures after three column osteotomies of the spine. Scoliosis Spinal Disord 2017;12:19. https://doi.org/10.1186/s13013-017-0127-x
crossref pmid pmc
16. Hu Z, Liu D, Zhu Z, Qiu Y, Liu Z. Using satellite rod technique in patients with severe kyphoscoliosis undergoing three-column osteotomy: a minimum of 2 years’ follow-up. Orthop Surg 2021;13:83–9. https://doi.org/10.1111/os.12836
crossref pmid pmc
17. Hallager DW, Gehrchen M, Dahl B, et al. Use of supplemental short pre-contoured accessory rods and cobalt chrome alloy posterior rods reduces primary rod strain and range of motion across the pedicle subtraction osteotomy level: an in vitro biomechanical study. Spine (Phila Pa 1976) 2016;41:E388–95. https://doi.org/10.1097/BRS.0000000000001282
crossref pmid
18. Luca A, Ottardi C, Sasso M, et al. Instrumentation failure following pedicle subtraction osteotomy: the role of rod material, diameter, and multi-rod constructs. Eur Spine J 2017;26:764–70. https://doi.org/10.1007/s00586-016-4859-8
crossref pmid
19. La Barbera L, Wilke HJ, Ruspi ML, et al. Load-sharing biomechanics of lumbar fixation and fusion with pedicle subtraction osteotomy. Sci Rep 2021;11:3595. https://doi.org/10.1038/s41598-021-83251-8
crossref pmid pmc
20. Berjano P, Xu M, Damilano M, et al. Supplementary delta-rod configurations provide superior stiffness and reduced rod stress compared to traditional multiplerod configurations after pedicle subtraction osteotomy: a finite element study. Eur Spine J 2019;28:2198–207. https://doi.org/10.1007/s00586-019-06012-2
crossref pmid
21. Gupta S, Eksi MS, Ames CP, et al. A novel 4-rod technique offers potential to reduce rod breakage and pseudarthrosis in pedicle subtraction osteotomies for adult spinal deformity correction. Oper Neurosurg 2018;14:449–56. https://doi.org/10.1093/ons/opx151
crossref pmid pmc
TOOLS
Share :
Facebook Twitter Linked In Google+ Line it
METRICS Graph View
  • 0 Crossref
  •   Scopus
  • 581 View
  • 19 Download
Related articles in ASJ

Biomechanical comparison of posterior short-segment fixation with or without intermediate screws for thoracolumbar burst fractures under normal and osteoporotic conditions: a finite element analysis2026 April;20(2)



ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
Department of Orthopedic Surgery, Asan Medical Center, University of Ulsan College of Medicine
88, Olympic-ro 43-gil, Songpa-gu, Seoul 05505, Korea
Tel: +82-2-3010-3530    Fax: +82-2-3010-8555    E-mail: asianspinejournal@gmail.com                
Korean Society of Spine Surgery
82, Gumi-ro 173beon-gil, Bundang-gu, Seongnam-si, Gyeonggi-do, 13620, Korea
Tel: +82-31-966-3413    Fax: +82-2-831-3414    E-mail: office@spine.or.kr                

Copyright © 2026 by Korean Society of Spine Surgery.

Developed in M2PI

Close layer
prev next