Dear Editor,
We thank the author of the Letter for the thoughtful comments on our study [
1]. We address the points concerning the multivariable analysis and the use of a single computed tomography (CT) slice below.
First, we would like to clarify the number of loosening events. Loosening was assessed at the screw level and occurred in 25 of 212 screws inserted in 47 patients, rather than in approximately five or six screws as stated in the Letter. The multivariable logistic regression included age, sex, screw trajectory Hounsfield unit (HU), screw length, and screw level. Screw diameter was not included.
Nevertheless, the number of events was limited, and the stability of the regression estimates warrants caution. The analysis also did not account for correlation among screws within the same patient. Its confidence intervals and p values should therefore be interpreted cautiously. The 123-HU cutoff was derived from receiver operating characteristic analysis of this cohort, not from the multivariable regression model. We agree that it should be considered an exploratory finding rather than a clinically validated threshold.
Second, we would like to clarify how the CT slice was selected. From the 1-mm-thick preoperative CT images used for screw planning, we selected an axial slice through the cephalocaudal center of the pedicle, parallel to the vertebral endplate. An elliptical region of interest was placed in the cancellous bone along the planned screw trajectory on that slice.
Our aim was to provide a practical assessment using routinely available preoperative CT images. This measurement does not assess bone quality throughout the three-dimensional screw trajectory. Trabecular density may vary along the path, and the final screw trajectory may differ from the planned trajectory. We assessed intraobserver and interobserver reproducibility of HU measurements using intraclass correlation coefficients. However, reproducibility of the measured HU values does not establish that a single slice represents the entire trajectory or that the same slice can be selected reproducibly across institutions.
Our findings show an association between HU values measured along the planned trajectory on a single slice and postoperative screw loosening. Larger studies should prespecify the method for slice selection and account for correlation among screws within patients, osteoporosis treatment, and construct-related factors. Comparison with volumetric HU assessment and external validation will help determine the generalizability of this practical method and the proposed 123-HU cutoff.
We appreciate the opportunity to clarify the purpose and limitations of our study.